Description
A very sensitive ammeter
Range | 10e-10 A - 10e-2 A full scale in 17 1* and 3* ranges |
---|---|
Accuracy | ±2% of full scale on 10e-2 to 10e-8 A ranges |
±4% of full scale on 3*10e-9 to 10e-10 A ranges. | |
Zero drift | Less than 0.5% of full scale per week |
Additional 0.05% per °C after a 10 min. warm-up | |
Input voltage drop | Less than 1 mV for full-scale meter deflection. |
Input resistance | Less than 0.1 Ohm on 10e-2 A range |
less than 10e7 Ohms on 10e-10 A range. | |
Max. input | Transient: 1000V for up to 3 secs. |
Continuous: 600V on 10e-10 to 10e-6 A ranges | |
Continuous: 12V on the 10e-2 A range. |
Note: This is a very sensitive instrument which will register a change in current if you pass your hand through your hair, etc..
Click here to go to the top of this page.