Universal Counter
Manufacturer: Hewlett Packard
Model: 5314 A
Description
100 MHz/100 ns Universal Counter. It features a seven-digit, seven-segment LED display
with overflow indication, seven function performance, and full input signal conditioning. Seven measurement
functions are: Frequency, Single-Shot Period, Period Average, Time Interval, Totalize, Ratio,
and Self Check. This is accomplished by a single LSI integrated circuit. The input signal is
AC coupled and can be conditioned as follows: slope selection, trigger level, and attenuation.
Specifications
INPUT Characteristics |
Range | Channel A (10 Hz - 100 MHz) |
| Channel B (10 Hz - 2.5 MHz) |
Sensitivity | 25 mV rms to 100/2.5 MHz. 75 mV p-p min. pulse with 5/50 ns (Channel A/Channel B). |
Impedance | 1 M Ohms NOMINAL shunted by < 30 pF. |
Attenuator | X1 or X20 NOMINAL (Channel A only) |
|
FREQUENCY (A) |
Range | 10 Hz to 10 MHz direct count |
| 1 MHz to 100 MHz prescaled by 10 |
Resolution | ± LSD |
Accuracy | ± LSD ± (time base error) X FREQ |
|
PERIOD (A) |
Range | 10 Hz to 2.5 MHz |
LSD displayed | (100 ns) / N for N=1 to 1000 |
Resolution | ± LSD ± 1.4 X (Trigger Error) / N |
Accuracy | Resolution ± (time base error) X PER |
|
TIME INTERVAL (A to B) |
Range | 250 ns to 1 s |
LSD displayed | 100 ns |
Resolution | ± LSD ± START Trigger Error ± STOP Trigger Error |
Accuracy | Resolution ± (time base error) X T1. |
LSD: Least Significant Digit.
Click here to go to the top of this page.
Operating Instructions
- Turn unit ON and set COM A/SEP switch to SEP position
- Connect input signal to INPUT A jack
- For frequency measurements, press FREQ A/START A IN. Make sure blue shift key is OUT. Set
SLOPE, ATTN and LEVEL A as desired and select appropriate resolution and frequencies.
- For period measurements, press PER A/T.I. A->B switch IN. Make sure blue shift key is OUT. Set SLOPE, ATTN and LEVEL and select desired sample size.
- For one-source time interval measurements, press PER A/T.I. A->B IN. Press the blue shift key IN. Set SLOPE, ATTN and LEVEL A/B controls and set resolution switch to N=1 mode.
- For two-source time interval measurements, follow procedure for one-source setup but connect start-time interval signal to INPUT A and stop-time interval signal to INPUT B.
- For ratio measurements, make sure FREQ A and PER A buttons are both OUT. Connect higher frequency signal to INPUT A and lower frequency signal to INPUT B. Adjust SLOPE, ATTN and LEVEL A/B controls and select desired sample size.
- For totalize measurements, press FREQ A IN and blue shift key IN. Set SLOPE, ATTN and LEVEL A and connect input signal to INPUT A (for freq. > 10 MHz, the 100 MHz/10 Hz resolution switch MUST be pressed (IN).