Description
Versatile meter capable of five different functions,
i.e. AC and DC currents, AC and DC voltages, and resistance.
Measured values are displayed on a digital readout. The
Hewlett-Packard meter is auto-ranging, i.e. the meter
automatically switches to the most appropriate range.
DC Voltage | |
---|---|
Ranges | ±0.11 V to 1100 V in 5 ranges (1000 V max.) |
Accuracy | 0.11 V ±(0.5% of reading + 0.2% of range) |
1.1 V, 11 V ±(0.5% of reading + 0.1% of range) | |
110 V, 1100 V ± (0.6% of reading + 0.1% of range) | |
Input Resistance | 10 Mohms ±5% |
DC Current | |
Ranges | ±0.11 A, 1.1 A |
Accuracy | ±(1.0% of reading + 0.2% of range) |
Impedance | 1 - 1.5 Ohms constant |
AC Voltage | |
Ranges | 0.11 Vrms to 1100 Vrms in 5 ranges (707 Vrms max) |
Accuracy | |
1.1 V to 1100 Vrms | ± (1.7% of reading + 0.5% of range), 45 Hz to 2 kHz |
0.11 Vrms | ±(1.7% of reading + 0.7% of range), 45 Hz to 2 kHz |
1.1 V to 1100 Vrms | ± (3.2% of reading + 0.7% of range), 2 kHz to 5 kHz |
0.11 Vrms | ± (5.2% of reading + 0.7% of range), 2 kHz to 5 kHz |
1.1 V to 1100 Vrms | ± (8.2% of reading + 1.1% of ranges), 5 kHz to 10 kHz |
0.11 Vrms | ± (18.2% of reading + 1.1% of range), 5 kHz to 10 kHz |
Input Resistance | 10 M Ohms ± 5% |
Input Capacitance | < 30 pF |
AC Current | |
Ranges | 0.11 A rms, 1.1 A rms. |
Accuracy | |
1.1 A rms | ±(2.2% of reading + 0.5% of range), 45 Hz to 2 kHz |
±(3.7% of reading + 0.7% of range) 2 kHz to 5 kHz | |
0.11 A rms | ±(2.2% of reading + 0.7% of range) 45 Hz to 2 kHz |
±(5.7% of reading + 0.7% of range) 2 kHz to 4 kHz | |
Impedance | 1 - 1.5 Ohms constant |
Resistance | |
Ranges | 1.1 kohm to 11000 kohm in 5 ranges |
Accuracy | |
110 k, 1100 kohm | ± (0.5% of reading + 0.1% of range) |
11000 k, 1.1 k, 11 kohm | ± (0.7% of reading + 0.1% of range) |
Click here to go to the top of this page.