Description
A versatile multimeter capable of measuring AC and DC
voltages and currents, and resistances. Note that AC
measurements are true root-mean-square values.
DC Voltage | |
---|---|
Accuracy | .05% on 20mV range |
.04% on 200 mV range | |
.03% on 2V through 200 V ranges | |
.035% on 1200 V range | |
Maximum input | 1200V DC + peak AC |
Input Resistance | 10 MOhm |
Sensitivity | 1 uV |
Response time | less than 0.7 sec. |
AC Voltage | |
Accuracy | 2%, 20Hz - 30Hz |
1%, 30Hz - 50Hz | |
0.3%, 50Hz - 20kHz | |
1%, 20kHz - 50kHz | |
2%, 50kHz - 100kHz | |
Maximum input | 600 Vdc or 1700 Vdc + peak ac |
Input impedance | 2 MOhm, less than 50 pF |
Sensitivity | 10 uV |
Response time | less than 4.5 sec. |
DC Current | |
Accuracy | .07%, 200 uA through 20 mA ranges |
.15%, 200 mA range | |
.5%, 2000 mA range | |
Maximum input | 2 A from less than 250 V source |
Sensitivity | 10 nA |
Voltage burden | less than 200 mV, 200 uA through 20 mA ranges |
less than 240 mV, 200 mA range | |
less than 600 mV, 2000 mA range | |
AC Current | |
Accuracy(200 uA thru 200 mA) | 2%, 20Hz - 30 Hz |
.9%, 30Hz - 10 kHz | |
Accuracy(2000 mA) | 2%, 20Hz - 30 Hz |
1.2%, 30Hz - 10 kHz | |
Maximum input | 2 A rms from less than 250 V rms source. |
Sensitivity | 10 nA |
Voltage burden | less than 220 mV, 200 uA through 20 mA ranges |
less than 240 mV, 200 mA range | |
less than 600 mV, 2000 mA range |
Resistance | |
---|---|
Accuracy (for 1 year at 23° ± 5° C) | 20, 200 Ohms ranges: ± (0.8% of reading) |
2, 20 and 200 k Ohms ranges: ± (0.03% of reading) | |
2 M Ohms range: ± (0.04% of reading) | |
20 M Ohms range: ± (0.15% of reading) | |
Sensitivity | 1 m Ohms on the 20 Ohms range |
Output Voltage | < 5 volts dc |