Precision magnetic field measuring instrument which provides advanced operating features. Critical performance parameters are monitored continually, and error messages are displayed whenever any of these parameters fall below preset standards.
Full Scale Ranges | 200 G, 2 kG, 20 kG, 20 mT, 200 mT, 2 T |
---|---|
Resolution | 0.1 G, 0.01 mT |
Accuracy | |
dc | ±2.0% of reading, including probe |
ac (45 Hz-100Hz) | ±2.5% of reading, including probe |
(100 Hz - 3 kHz) | ±2.0% of reading, including probe |
(3 kHz - 5 kHz) | ±6.0% of reading, including probe (useable with reduced accuracy to 12 kHz) |